Other Titles

A health system change in practice in the care of patients with tracheal T tube during a code

Abstract

Session presented on Friday, September 26, 2014:

Purpose: Implement practice change across the healthsystem on the management of patients with a tracheal T Tube during a code.

Methods: The absence of a universal adapter for a tracheal T Tube, in the unit, which is needed to fit and seal onto the ambag during a code, resulted in a catastrophic event, which rendered the patient brain anoxic with resultant persistent and continuous vegetative state. The head and neck units have specific guidelines and standards for the care of a patient with a tracheal T Tube. Included in its onboarding of new staff are very specific staff competencies in suctioning and airway management in the event of a code on a patient with a tracheal T Tube. An algorithm was agreed upon by the departments of anesthesia, nursing and head and neck in managing such patients. These competencies specifically state that a universal adapter be provided by Respiratory Therapy. In this particular code, per guideline, Nursing requested an adapter but Respiratory Therapist on duty stated he has no knowledge of a tracheal T tube, nor is he aware of an adapter. In addition, the code blue team was not aware that the end of the endotracheal tube could be used as an adapter to connect to the tracheal T tube to maintain a seal and bag the patient. Instead, the anesthesiologist plugged the tracheal T tube and attempted to intubate the patient orally. The algorithm was not followed in this case, which calls for the removal of the tracheal T Tube and the insertion of a regular tracheostomy tube. In the absence of the adapter, nursing staff aggressively called for the removal of the T tube and the placement of a regular tracheostomy tube to no avail. The patient sustained a catastrophic brain anoxia with resultant vegetative state. We have had similar events in the head and neck service where code blue staff were unfamiliar with the handling of specialty patients with surgically implanted airways. Three previous codes were root caused analysed, which resulted in the interprofessional and interdisiplinary adoption of an algorithm. Nursing led this effort and the departments of anesthesia and head and neck chairs and their residency education chairs were involved. Anesthesia and head and neck residents and nursing staff were educated on the difficult intubation algorithm. The algorithm is now placed in all emergency crash carts. The goal was to prevent incidents like the one described.

Results: There is definite knowledge deficits on respirary therapy and code blue MDs on the management of patients with tracheal T Tubes. The health system failed in that the adapter was not available when needed. The adapter should have been made available at the unit level, despite its cost or classification as a slow moving item. These adapters should have been listed in our PAR and placed at our supply room. The ET tube connector could have been used as a substitute but nurses and MDs did not utilize it. Massive re-education of RNS, RTs,MDs , inclusion of tracheal T tube adapters in competencies, review of difficult intubation algorithm, presence of T Tube adapters in unit emergency crash cart, non dependence on RT staff, empowering head and neck RNs to oblige code blue staff to follow set algorithm, specific orders and pictures of tracheal T tube on airway management in EHR were instituted.

Conclusion: To date, no repeat event as described above has occurred. To sustain this result, the head and neck Clinical Nurse Specialist meets every quarter with the Anesthesia Residency Education Chair to review the difficult intubation algorithm to ensure that code blue anesthesiologists are competent in a code involving a head and neck patient with a tracheal T Tube.

Author Details

Malou Blanco-Yarosh, MSN

Sigma Membership

Non-member

Lead Author Affiliation

University of California, Los Angeles Health, Los Angeles, California, USA

Type

Presentation

Format Type

Text-based Document

Study Design/Type

N/A

Research Approach

N/A

Keywords:

Transdisciplinary Education, Transdisciplinary Change in Practice, Tracheal T Tube

Conference Name

Leadership Summit 2014

Conference Host

Sigma Theta Tau International

Conference Location

Indianapolis, Indiana, USA

Conference Year

2014

download (267 kB)

Rights Holder

All rights reserved by the author(s) and/or publisher(s) listed in this item record unless relinquished in whole or part by a rights notation or a Creative Commons License present in this item record.

All permission requests should be directed accordingly and not to the Sigma Repository.

All submitting authors or publishers have affirmed that when using material in their work where they do not own copyright, they have obtained permission of the copyright holder prior to submission and the rights holder has been acknowledged as necessary.

Review Type

Abstract Review Only: Reviewed by Event Host

Acquisition

Proxy-submission

Share

COinS
 

A health system change in practice in the care of patients with tracheal T tube during a code

Indianapolis, Indiana, USA

Session presented on Friday, September 26, 2014:

Purpose: Implement practice change across the healthsystem on the management of patients with a tracheal T Tube during a code.

Methods: The absence of a universal adapter for a tracheal T Tube, in the unit, which is needed to fit and seal onto the ambag during a code, resulted in a catastrophic event, which rendered the patient brain anoxic with resultant persistent and continuous vegetative state. The head and neck units have specific guidelines and standards for the care of a patient with a tracheal T Tube. Included in its onboarding of new staff are very specific staff competencies in suctioning and airway management in the event of a code on a patient with a tracheal T Tube. An algorithm was agreed upon by the departments of anesthesia, nursing and head and neck in managing such patients. These competencies specifically state that a universal adapter be provided by Respiratory Therapy. In this particular code, per guideline, Nursing requested an adapter but Respiratory Therapist on duty stated he has no knowledge of a tracheal T tube, nor is he aware of an adapter. In addition, the code blue team was not aware that the end of the endotracheal tube could be used as an adapter to connect to the tracheal T tube to maintain a seal and bag the patient. Instead, the anesthesiologist plugged the tracheal T tube and attempted to intubate the patient orally. The algorithm was not followed in this case, which calls for the removal of the tracheal T Tube and the insertion of a regular tracheostomy tube. In the absence of the adapter, nursing staff aggressively called for the removal of the T tube and the placement of a regular tracheostomy tube to no avail. The patient sustained a catastrophic brain anoxia with resultant vegetative state. We have had similar events in the head and neck service where code blue staff were unfamiliar with the handling of specialty patients with surgically implanted airways. Three previous codes were root caused analysed, which resulted in the interprofessional and interdisiplinary adoption of an algorithm. Nursing led this effort and the departments of anesthesia and head and neck chairs and their residency education chairs were involved. Anesthesia and head and neck residents and nursing staff were educated on the difficult intubation algorithm. The algorithm is now placed in all emergency crash carts. The goal was to prevent incidents like the one described.

Results: There is definite knowledge deficits on respirary therapy and code blue MDs on the management of patients with tracheal T Tubes. The health system failed in that the adapter was not available when needed. The adapter should have been made available at the unit level, despite its cost or classification as a slow moving item. These adapters should have been listed in our PAR and placed at our supply room. The ET tube connector could have been used as a substitute but nurses and MDs did not utilize it. Massive re-education of RNS, RTs,MDs , inclusion of tracheal T tube adapters in competencies, review of difficult intubation algorithm, presence of T Tube adapters in unit emergency crash cart, non dependence on RT staff, empowering head and neck RNs to oblige code blue staff to follow set algorithm, specific orders and pictures of tracheal T tube on airway management in EHR were instituted.

Conclusion: To date, no repeat event as described above has occurred. To sustain this result, the head and neck Clinical Nurse Specialist meets every quarter with the Anesthesia Residency Education Chair to review the difficult intubation algorithm to ensure that code blue anesthesiologists are competent in a code involving a head and neck patient with a tracheal T Tube.